• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. 20" CENTURY STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. 20" CENTURY STERILIZER Back to Search Results
Device Problems Fire (1245); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the sterilizer and found the root cause of the reported event to be a failure in the water pump of the steam generator.As the water pump was not operating properly, this allowed the electric motor to stop rotating and overheat subsequently causing the reported event to occur.The wire insulation within the electric motor had evidence of charring.The unit was installed in 2005 making it approximately 14 years old.The technician replaced the water pump and the electric motor, tested the unit, confirmed it to be operating according to specification, and returned it to service.No additional issues have been reported.
 
Event Description
The user facility reported that fire and smoke were observed after opening the back of their 20" century sterilizer.User facility personnel extinguished the flames and dispatched the fire department.No evacuations were conducted.No report of injury or inhalation/irritation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20" CENTURY STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key8457495
MDR Text Key140124120
Report Number3005899764-2019-00027
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-