| Catalog Number |
168-600 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a right side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified yellow and brown particles in the device's outer surface and one linear opening.A leak test was performed which identified an opening.A microscopic analysis was performed which identified one sharp opening.A dimension measurement in the shell was performed which identified the shell thickness within specification.The fill inspection was performed and identified no blockage in the valve.Based on the device analysis, the final assessment is one sharp opening assessed as unidentified tear opening.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a right side deflation.Device has been explanted.
|
| |
|