Catalog Number 6260-4-090
Medical Device Problem Codes Crack (1135); Fracture (1260); Misassembled (1398)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not available.
 
Event or Problem Description
While performing a revision of patient's hip, the impactor adapter wasn't fully screwed into the stem.When the instrument was back slapped, the threads sheared off in the stem.Another device was immediately available and removal was completed with no surgical delay.Rep reported that the surgeon and hospital have no further information available.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a gmrs adaptor was reported.The event was confirmed after review of the provided photograph.Method & results product evaluation and results: the reported device was not returned however a photograph was provided for review.Review of the photograph noted the following: the threaded end of the device has fractured just below the first thread.The device shows some staining from biological material but is otherwise, unremarkable.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as the operative report as well as device return are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
While performing a revision of patient's hip, the impactor adapter wasn't fully screwed into the stem.When the instrument was back slapped, the threads sheared off in the stem.Another device was immediately available and removal was completed with no surgical delay.Rep reported that the surgeon and hospital have no further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCREYNOLDS IMPACTOR ADAPTER
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8457823
Report Number0002249697-2019-01462
Device Sequence Number16875399
Product Code LPH
UDI-Device Identifier07613327217896
UDI-Public07613327217896
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K153345
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6260-4-090
Device Lot NumberTACKA0E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2019
Supplement Date Received by Manufacturer04/12/2019
Initial Report FDA Received Date03/27/2019
Supplement Report FDA Received Date05/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-