| Catalog Number |
168-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/20/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified white/yellow particles on the device's outer surface and one curved opening.A leak test was performed which identified one opening.A microscopic analysis was performed which identified one sharp opening.A dimension measurement in the shell was performed which identified the thickness within specification.A fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is one sharp opening assessed as an unidentified (tear) opening.Reason for reoperation due to deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side deflation.Device has been explanted.
|
| |
|