| Catalog Number |
4000-07 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Failure to Anastomose (1028)
|
| Date of Event |
02/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We have received the complaint device for evaluation and we have confirmed the reported incident.The gap distance between clips did not consistently pass our specification of 0.006".The gap between the clip cover and the applier jaw was within our acceptable limit and there was adequate stack force being applied on the clips.At this time, we are inconclusive about the root cause of the defect but the investigation is on-going.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.We have not received any other complaints of a similar nature for devices from this lot from other customers.There was no injury to the patient.The procedure was completed by using another anastoclip vcs device ( lot# unknown ).Please note that we have also submitted manufacturer's report number: 1220948-2019-00031 for another case that was reported to us by the same surgeon.
|
| |
|
Event or Problem Description
|
|
One large anastoclip vcs appliers failed to close the vascular tissue properly during anastomosis.This is report 2 of 2.
|
| |
|