(b)(4).Batch # r92008.Device analysis: the analysis results found that the mcm30 device was returned with a clip in the jaws.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining 21 clips as intended.As the device was found to be fully functional, it could not be determined what may have caused the reported incident.No conclusion could be reached as to what may have caused the reported incident.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch r92008 number, and no non-conformances were identified.
|