Model Number CM2010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported during the sensor implant procedure, the guidewire snapped inside the swan while the physician was trying to advance into the implant site.The issue was observed prior to the introduction of the cardiomems delivery system.Reportedly, the physician was able to snare the two pieced out and successfully implant the sensor and calibrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOMEMS GUIDEWIRE
Common Device Name
SYSTEM, HEMODYNAMIC, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. - CARDIOMEMS
387 technology circle nw suite 500
atlanta GA 30313
MDR Report Key8460725
Report Number3004936110-2019-00163
Device Sequence Number11484860
Product Code MOM
UDI-Device Identifier15414734509562
UDI-Public15414734509562
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/24/2020
Device Model NumberCM2010
Device Catalogue NumberCM2010
Device Lot NumberMO43675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/19/2019
Supplement Date Received by Manufacturer06/20/2019
03/31/2020
Initial Report FDA Received Date03/28/2019
Supplement Report FDA Received Date06/21/2019
03/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight50
-
-