Catalog Number 1012449-08
Medical Device Problem Codes Loose or Intermittent Connection (1371); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Date of event - estimated.Evaluation summary: visual and functional analysis were performed on the returned device.The reported loose or intermittent connection was not confirmed; however, the device was returned with a hypotube separation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that may have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty or for the noted hypotube separation.There is no indication of a product quality issue with respects to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the connection of the indeflator came apart from the shaft of the 3.0 x 8 mm nc trek balloon dilatation catheter.Another balloon catheter was used to complete the procedure with no reported adverse patient effect or clinically significant delay.Device analysis revealed the hypotube was separated at the distal end of the hub.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8461968
Report Number2024168-2019-02405
Device Sequence Number2839731
Product Code LOX
UDI-Device Identifier08717648151859
UDI-Public08717648151859
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number1012449-08
Device Lot Number81120G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2019
Initial Report FDA Received Date03/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-