MAUDE Adverse Event Report: INSET¿30; INSET 30 60/13 GREY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSET¿30; INSET 30 60/13 GREY Back to Search Results
Lot Number 5076075
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Hyperglycemia (1905)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in (b)(6).It was reported that the patient had cracked/broken tubing.The patient reported that during the night the tubing broke and separated from the needle.The blood glucose was greater than or equal to 250 mg/dl but less than 500 mg/dl.Upon investigation, it was found that the tubing was detached from the tubing connector (due to insufficient glue in the connection).No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET¿30
Common Device Name
INSET 30 60/13 GREY
MDR Report Key8462874
Report Number3003442380-2019-00786
Device Sequence Number16465802
Product Code FPA
UDI-Device Identifier05705244007109
UDI-Public05705244007109
Combination Product (Y/N)N
PMA/510(K) Number
K061374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date02/01/2019
Device Lot Number5076075
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2016
Initial Report FDA Received Date03/28/2019
Patient Sequence Number1
-
-