• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG IN-SITU PLATE CUTTER, UPPERFACE/MIDFACE; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG IN-SITU PLATE CUTTER, UPPERFACE/MIDFACE; INSTRUMENT Back to Search Results
Catalog Number 62-18330
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not yet been received at the manufacturer for testing.An evaluation will be conducted upon receipt of the device, and a follow-up report will be submitted after the quality investigation is complete.
 
Event or Problem Description
It was reported that the in-situ cutter fractured on the rivet hole during a craniotomy case.The cutter was used to cut 1.7mm plate.
 
Additional Manufacturer Narrative
The investigation shows that the product subject to the complaint had reached its life time at the event date of the reported incident.The relevant quality documents indicate that the product was manufactured according to specifications and was accepted into stock without any discrepancies.There is no indication for a quality issue with the article in question, nor is there any indication for a deviation from the defined quality specifications within development and or manufacturing.No indications for unusual or unexpected circumstances could be identified.Moreover, it is to be concluded that the reported incident originated from a common case of wear and tear resulting from aging as it is to be expected with mechanical medical devices subject to reprocessing after reaching their life time, not making any further action necessary.This complaint will be considered in statistical analyses at frequent intervals.
 
Event or Problem Description
It was reported that the in-situ cutter fractured on the rivet hole during a craniotomy case.The cutter was used to cut 1.7mm plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN-SITU PLATE CUTTER, UPPERFACE/MIDFACE
Common Device Name
INSTRUMENT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
MDR Report Key8464852
Report Number0008010177-2019-00014
Device Sequence Number11038844
Product Code LXH
UDI-Device Identifier04546540366719
UDI-Public04546540366719
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number62-18330
Device Lot NumberUNKNOWN/FROM STOCKED CONSIGNME
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Initial Date Received by Manufacturer 03/25/2019
Supplement Date Received by Manufacturer03/25/2019
Initial Report FDA Received Date03/29/2019
Supplement Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-