| Catalog Number |
B15LT |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # r94a1l.Device analysis: the analysis results found that the b15lt device was returned with the seal damaged and torn; the torn piece of the seal was returned.The device was visually inspected in order to evaluate the condition of the seal and the damage was confirmed and the separate piece of the seal was observed to have a mark which suggests that a pointy instrument was inserted through the trocar with excessive force.Per instructions for use: "use caution when introducing or removing instruments through the trocar sleeve in order to prevent inadvertent damage to the seals which could result in loss of pneumoperitoneum.Special care should be used when inserting sharp or angled edged endoscopic instruments to prevent tearing the seal." a manufacturing record evaluation was performed for the finished device batch/lot number, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic prostatectomy surgery, endopath xcel trocar universal seal was broken and one piece of this seal was found separately from the operation site.Dark gray leaflet on the inside of the universal seal become broken.This piece of seal was removed, and surgery continued in usual manner.There were no patient consequences.
|
| |
|