• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COYOTE ES; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COYOTE ES; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24691
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/07/2019
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event Description
It was reported that catheter entrapment occurred.The 99% stenosed target lesion was located in a moderately calcified and mildly tortuous vessel below the knee.After a non- bsc guide wire crossed the lesion, a 2mm x 40mm x 145cm coyote es balloon catheter was advanced for pre-dilatation.However, the balloon catheter did not advance at the area before the lesion, and it became stuck with the wire.The device was remove together as a whole system.The procedure was completed with a non-bsc device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE ES
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8464935
MDR Text Key140333338
Report Number2134265-2019-03180
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08714729767190
UDI-Public08714729767190
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K093636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 03/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/17/2021
Device Model Number24691
Device Catalogue Number24691
Device Lot Number0022136128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-