WRIGHTS LANE SYNTHES USA PRODUCTS LLC SYNFIX® EVOLUTION SCREWDRIVER W/O THREAD LOCK SLEEVE; SCREWDRIVERS Back to Search Results
Model Number 03.835.013
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2019, the patient underwent anterior lumbar interbody fusion (alif) l5-s1.During the surgery, the surgeon was screwing in the screws and the tip of the screwdriver broke off.There were no fragments generated from the broken device.The surgery was successfully completed with no surgical delay.Patient outcome is unknown.Concomitant medical products reported: unknown screws (part#unknown, lot#unknown, quantity unknown).This report is for one (1) synfix® evolution screwdriver w/o thread lock sleeve.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A device history record (dhr) review was conducted: part: 03.835.013.Lot: l576088.Manufacturing site: haegendorf.Release to warehouse date: 27.Oct.2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Only top level of the device history record reviewed as sub-components are not lot tracked.A product investigation was conducted.Visual inspection: the synfix® evolution screwdriver w/o thread lock sleeve (part # 03.835.013, lot # l576088, mfg # 27-oct-2017) was received at us cq with distal joint very loose.This is not consistent with the reported complaint condition.Therefore, the complaint is not confirmed.Dimensional inspection: dimensional analysis was not performed as there is no relevant dimension to the distal joint being loose.Document/specification review: the relevant drawing(s) was reviewed; conclusion: there is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces.No new malfunctions were observed during the course of this investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: correction-initial reporter phone number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFIX® EVOLUTION SCREWDRIVER W/O THREAD LOCK SLEEVE
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8465091
Report Number2939274-2019-57177
Device Sequence Number240312
Product Code HXX
UDI-Device Identifier10705034816234
UDI-Public(01)10705034816234
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.835.013
Device Catalogue Number03.835.013
Device Lot NumberL576088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2019
Initial Date Received by Manufacturer 03/05/2019
Supplement Date Received by Manufacturer03/05/2019
04/15/2019
Initial Report FDA Received Date03/29/2019
Supplement Report FDA Received Date03/29/2019
04/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
-
-