• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH CITATION TMZF HA STEM #6 RIGHT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH CITATION TMZF HA STEM #6 RIGHT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6265-5116
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.When completed, the evaluation summary will be submitted in a supplemental report.
 
Event or Problem Description
Surgeon implanted the femoral stem into the canal and noticed there were some significant scratches on the trunnion.Due to the use of the use of a ceramic head he opted to remove the stem and implant a new one.Surgeon says that it was scratched out of the package however the stem was implanted before the scratches were noticed so there is no way of knowing.
 
Event or Problem Description
Surgeon implanted the femoral stem into the canal and noticed there were some significant scratches on the trunnion.Due to the use of the use of a ceramic head he opted to remove the stem and implant a new one.Surgeon says that it was scratched out of the package however the stem was implanted before the scratches were noticed so there is no way of knowing.
 
Additional Manufacturer Narrative
An event regarding damage involving a citation stem was reported.The event was confirmed by device images.Method & results: product evaluation and results: damage observed on the stem consistent with attempted implantation/explantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: not performed as medical records were not provided for review.Product history review: a device history review confirmed all devices accepted into finished goods conformed to specification.Complaint history review: no other events were reported for the lot indicated.Conclusions: the investigation confirmed the reported event of damage on the implant however, the exact root cause could not be determined.Further assessment beyond an examination with an engineer could not be performed as the damage visible on the trunnion is consistent with implantation and explantation damage.Further information is needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CITATION TMZF HA STEM #6 RIGHT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8466807
Report Number0002249697-2019-01480
Device Sequence Number16835644
Product Code LPH
UDI-Device Identifier07613153074397
UDI-Public07613153074397
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue Number6265-5116
Device Lot Number52420502
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2019
Initial Date Received by Manufacturer 03/01/2019
Supplement Date Received by Manufacturer04/18/2019
Initial Report FDA Received Date03/29/2019
Supplement Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
-
-