NEODENT - JJGC S.A. CM DRIVE ACQUA IMPLANT 5.0X13; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.686
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 09/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
 
Event or Problem Description
The clinician reported 9.5 months after the dental implant and abutment were placed in ada site 7 in the mouth, the implant did not achieve osseointegration in type ii bone.The implant was torqued to 50 ncm.The implant will be forwarded to the manufacturer for investigation.There were no reported patient complications.
 
Event or Problem Description
The clinician reported 9.5 months after the dental implant and abutment were placed in ada site 7 in the mouth, the implant did not achieve osseointegration in type ii bone.The implant was torqued to 50 ncm.The implant will be forwarded to the manufacturer for investigation.There were no reported patient complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system that is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a new follow-up report will be send.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE ACQUA IMPLANT 5.0X13
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key8467054
Report Number3008261720-2019-01318
Device Sequence Number240198
Product Code DZE
UDI-Device Identifier07899878016878
UDI-Public07899878016878
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/29/2019
Report Date (Section F) 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/09/2018
Device Catalogue Number140.686
Device Lot Number800222093I
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/05/2019
Device Age1 MO
Event Location Other
Date Report to Manufacturer03/06/2019
Initial Date Received by Manufacturer 03/06/2019
Supplement Date Received by Manufacturer06/21/2019
Initial Report FDA Received Date03/29/2019
Supplement Report FDA Received Date07/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABUTMENT REMOVAL; ABUTMENT REMOVAL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight88
-
-