MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523RNAL
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 180 mg/dl at the time of the incident.During seating the force sensor did not detect the reservoir.The customer stated that the drive support cap was normal.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device was expected to be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump passed the displacement test.Device received compromised forced sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test due to compromised forced sensor system alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8467868
Report Number3004209178-2019-74682
Device Sequence Number16479147
Product Code OYC
UDI-Device Identifier00643169476486
UDI-Public(01)00643169476486
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523RNAL
Device Catalogue NumberMMT-523RNAL
Device Lot NumberA0523RNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Initial Date Received by Manufacturer 03/26/2019
Supplement Date Received by Manufacturer05/21/2019
Initial Report FDA Received Date03/30/2019
Supplement Report FDA Received Date05/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient Weight59
-
-