INSET¿; INSET I 60/6 GREY Back to Search Results
Lot Number 5153787
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.It was reported that the patient had a tubing issue as it was cracked/broken.The tubing came apart from the infusion set prior to use.In the investigation it was found that the glue was missing from the tubing connector (glue was misplaced on the tubing connector).No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET¿
Common Device Name
INSET I 60/6 GREY
MDR Report Key8467966
Report Number3003442380-2019-00802
Device Sequence Number13831996
Product Code FPA
UDI-Device Identifier05705244006874
UDI-Public05705244006874
Combination Product (Y/N)N
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/01/2019
Device Lot Number5153787
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2017
Initial Report FDA Received Date03/30/2019
Patient Sequence Number1
-
-