• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIO¿; INSET II 46/6 PCC BLUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIO¿; INSET II 46/6 PCC BLUE Back to Search Results
Lot Number 5152541
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient's mother reported that the tubing kept coming out of its clip and detachment occurred where the clear blue piece came out of blue housing while the patient was at school.This was the third time that the tubing came off.The site location was at the bottom and the pump was located on the hip/belt.The infusion was in use for couple of hours.Reportedly, the infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to the patient's body.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIO¿
Common Device Name
INSET II 46/6 PCC BLUE
MDR Report Key8467971
Report Number3003442380-2019-00854
Device Sequence Number1501317
Product Code FPA
UDI-Device Identifier05705244007314
UDI-Public05705244007314
Combination Product (Y/N)N
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Lot Number5152541
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2017
Initial Report FDA Received Date03/31/2019
Patient Sequence Number1
-
-