• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIO¿; INSET II 80/6 PCC GREY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIO¿; INSET II 80/6 PCC GREY Back to Search Results
Lot Number 5183330
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient reported that his infusion set was due to be changed when he was about to pull back the serter to lock it in place during insertion and the tubing detached by the quick release while it was still in the serter.So, he had to start over again with a new set.The infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to the patient's body.Upon investigation, it was found that the tubing was detached from the tubing connector due to missing glue.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIO¿
Common Device Name
INSET II 80/6 PCC GREY
MDR Report Key8468004
Report Number3003442380-2019-00735
Device Sequence Number16510415
Product Code FPA
UDI-Device Identifier05705244007796
UDI-Public05705244007796
Combination Product (Y/N)N
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date03/01/2020
Device Lot Number5183330
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2017
Initial Report FDA Received Date03/31/2019
Patient Sequence Number1
-
-