MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PREP IM ENCHANCE TOTAL HIP KIT; CEMENT OBTURATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PREP IM ENCHANCE TOTAL HIP KIT; CEMENT OBTURATOR Back to Search Results
Catalog Number 121010
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to infection.Primary surgery was performed on (b)(6) 2019.Patient left hip wound would not heal and was infected.Patient had a washout and debridement prior to first stage revision on (b)(6) 2019, acetabular implants and femoral component were removed.Antibiotic cement spacer implanted.
 
Additional Manufacturer Narrative
It was reported that a revision surgery was performed due to infection.The complained devices have not been returned for investigation and the reported failure mode could therefore not independently be confirmed.No clinical supporting documents have been provided for the inclusion in the medical assessment.A thorough medical assessment could therefore not be conducted.Three out of five batch numbers in scope have been communicated.No deviations from the standard manufacturing process were revealed for these batch records.Sterilization was conducted according to procedure and within specification.Based on available information the root cause for the reported infection cannot be identified and stays undetermined.The need for corrective action is not indicated.Nevertheless, smith and nephew will continue to monitor the devices for similar issues.The complaint will be reopened should additional information or the complained devices be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREP IM ENCHANCE TOTAL HIP KIT
Common Device Name
CEMENT OBTURATOR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8468041
Report Number1020279-2019-01270
Device Sequence Number16538017
Product Code LZN
UDI-Device Identifier03596010457127
UDI-Public03596010457127
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K791125
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121010
Device Lot Number18GSM0212
Initial Date Received by Manufacturer 03/07/2019
Supplement Date Received by Manufacturer03/07/2019
Initial Report FDA Received Date03/31/2019
Supplement Report FDA Received Date06/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-