Catalog Number 71338560
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
A revision surgery was performed due to dislocation.
 
Additional Manufacturer Narrative
The associated r3 20 deg 4 xlpe acetabular liner and oxonium femoral head were not returned for evaluation.Therefore a product analysis could not be performed.However, device details were provided for the liner.Thus a review of the manufacturing records and complaint history was conducted.The review of the manufacturing records for the liner did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the part revealed no prior complaints for the listed failure mode with the same batch number.After repeated requests, smith and nephew has been unable to obtain batch information of the femoral head.There is no information that would suggest the implanted devices failed to meet specifications.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.Our clinical evaluation noted that the x-rays provided show the cup at a steep inclination angle of approx 55 degrees which could be a reason for the recurrent dislocations.Details regarding the patient¿s weight-bearing status, medical history, bone quality, fall/trauma history, and other additional clinical relevant information have not been provided.Based on the limited information provided and without the return of the device for evaluation the root cause of the reported cup shifting and dislocations cannot be confirmed but the cup position is the likely cause.The future impact to the patient beyond the revision cannot be determined.Without the return of the actual product involved and no batch information available for the femoral head, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should the devices or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG +4 XLPE ACET LNR 36MM X 60MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8468064
Report Number1020279-2019-01280
Device Sequence Number11155863
Product Code MBL
UDI-Device Identifier03596010598752
UDI-Public03596010598752
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71338560
Device Lot Number09EM15130
Initial Date Received by Manufacturer 03/07/2019
Supplement Date Received by Manufacturer03/07/2019
Initial Report FDA Received Date03/31/2019
Supplement Report FDA Received Date07/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age86 YR
-
-