The associated sureshot targeter was returned and evaluated.A visual inspection of the device noted small cuts on the cord.A functional evaluation was performed by connecting the sureshot targeter to the sureshot interface unit which could not confirm the stated failure.The sureshot targeter has been recognized by the unit.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.The device was manufactured in 2015, which suggests it has been in use for some time.The sureshot targeter is a reusable device that can be exposed to numerous surgeries; damage from repeated use can occur.Several potential factors that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to/after each use and cleaning.A second generation targeter has been released and is available for use.Please reference device 71692851 when replenishing.No additional actions are being taken at this time; however smith and nephew will continue to monitor for future complaints and investigate further as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
|