• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA AG INVIA MOTION+ ENDURE SINGLE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA AG INVIA MOTION+ ENDURE SINGLE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 0874010
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was requested for return and a replacement device was sent to the customer.An evaluation of the device confirmed that this complaint is related to an issue in which otherwise fully functioning pumps display a battery missing error notification, which can be caused by repetitive unplugging and plugging when the battery is fully charged.This situation can potentially lead to a therapy interruption.The user is notified of such a malfunction via an audible and visual notification and is instructed via the instructions for use to contact their healthcare provider in order to receive a replacement device.
 
Event or Problem Description
On (b)(6) 2018, the customer alleged to medela llc that the motion negative pressure wound therapy device was not holding a charge while on a patient; however, there was no serious injury, the patient did not experience wound deterioration and the patient did not require consultation and/or treatment by a medical professional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIA MOTION+ ENDURE SINGLE
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
MEDELA AG
lattichstrasse 4b
baar zug, 6341
SZ  6341
MDR Report Key8469345
Report Number1419937-2019-00052
Device Sequence Number1846220
Product Code OMP
UDI-Device Identifier07612367052849
UDI-Public07612367052849
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K161128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0874010
Device Catalogue Number0874010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2019
Initial Report FDA Received Date04/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-