Lot Number 5220766
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient reported tubing detachment and stated that while she was changing her set, she put it on her body and the end popped out.The infusion set disconnected and the piece that was in her body popped off at the site.Patient's weight was (b)(6).Upon investigation, it was found that the tubing was detached from the tubing connector due to missing glue.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURE-T® PARADIGM®
Common Device Name
CONTACT DETACH G29 60/8 PCC
MDR Report Key8469611
Report Number3003442380-2019-00589
Device Sequence Number19190469
Product Code FPA
UDI-Device Identifier05705244005792
UDI-Public05705244005792
Combination Product (Y/N)N
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/01/2021
Device Lot Number5220766
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2018
Initial Report FDA Received Date04/01/2019
Patient Sequence Number1
-
-