Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.It came with no electrode.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.Visual inspection found no damage to the device.The device was plugged into a 40k ohm test box and activated in both the cut and coag mode with satisfactory results.The device plugged into a generator and activated normally.The device passed hipot testing, showing that there was no unexpected electrical leakages.The investigation did not identify any defects with the pencil that could have contributed to the reported event.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, confirm proper electrosurgical generator power setting before proceeding with surgery.Use the lowest power setting to achieve the desired surgical effect.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|