| Model Number |
24695 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr.: the device was returned for evaluation.The device returned with its original pouch, the pouch was opened and the device was out of it.A visual inspection on the device revealed that there was a small piece of a foreign matter that looks like plastic within the device, the device was not cracked.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that there was a crack on the device.A 10pk gateway plus was selected for use.During the preparation, it was noted that there was a crack on the device.The procedure was completed with another of the same device.There were no patient complications reported.However,a analysis revealed that there was a small piece of a foreign matter that looks like plastic and it caused an optical effect that seemed to be a crack in the device material.
|
| |
|