Catalog Number 508-11-58F
Medical Device Problem Codes Positioning Problem (3009); Migration (4003)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported that the patient's left hip was revised due to the shell shifting and becoming anteverted.Surgeon confirmed shifting by comparing post-op x-rays from primary implant on (b)(6) 2019 and post-op x-rays in (b)(6) 2019.Possible cause or contributor to shell shifting was not reported to rep.A shell, 2 screws, liner, biolox ceramic head and accolade ii stem were revised to a trident tritanium revision shell with 5 screws, an adm/mdm liner construct, lfit v40 metal head, and accolade ii stem.There are no allegations against the liner, head, or stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL MULTI
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8470309
Report Number0002249697-2019-01491
Device Sequence Number1664689
Product Code LPH
UDI-Device Identifier04546540432230
UDI-Public04546540432230
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Catalogue Number508-11-58F
Device Lot Number65988601
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2019
Initial Report FDA Received Date04/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-