Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 01/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Concomitant medical products: therapy date: (b)(6) 2019.Initial reporter also sent report to fda: the user facility is foreign; therefore a facility medwatch report will not be available.Report source - (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00101 & 0001032347-2019-00192.
 
Event or Problem Description
It was reported an exploratory surgery to determine whether the patient has developed an infection was performed.No infection was confirmed from either tissue or fossa component.The surgeon found that there was a communicating channel between the ear canal and the tmj area.The patient has repeatedly suffered for many years from ear infections so issues with the joint area most likely due to this.The fossa was removed during this procedure.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of evidence that the revision was performed.It was later reported that there was "no infection so far confirmed from either tissue or fossa component." the surgeon also indicated he "found that there was a communicating channel between the ear canal and the tmj area.Patient has repeatedly suffered for many years from ear infections so issues with the joint area most likely due to this." device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Investigation results concluded that the reported event was due to patient condition.There was nothing to indicate that there was an alleged malfunction of the implants.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00192-1 & 0001032347-2019-00101-2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ SYSTEM CROSS DRIVE SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key8470513
Report Number0001032347-2019-00191
Device Sequence Number13899254
Product Code JEY
UDI-Device Identifier00841036057459
UDI-Public00841036057459
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K910038
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number99-6579
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/10/2019
Supplement Date Received by Manufacturer05/30/2019
Initial Report FDA Received Date04/01/2019
Supplement Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age24 YR
-
-