CORDIS CORPORATION POWERFLEXPRO 8MM2CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400802S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot 17731743 presented no issues during the manufacturing process that can be related to the reported event.The device was not returned for analysis.A device history record (dhr) review was performed and showed that these lots of products met all the requirements per the applicable manufacturing quality plan.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, a power flex pro was used for a post dilation after a stent was implanted, however, it ruptured within its nominal pressure.There was no reported patient injury.It was replaced with a new balloon catheter and the procedure was continued.The lesion was the iliac artery.
 
Additional Manufacturer Narrative
Corrected data: initial reporter name and address.As reported, a powerflex pro was used for a post dilation after a stent was implanted; however, it ruptured within its nominal pressure.There was no reported patient injury.The lesion was the iliac artery.It was replaced with a new balloon catheter and the procedure was continued.The product was not returned for analysis.A product history record (phr) review of lot 17731743 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.With the information available and without return of the product for analysis, it is not possible to draw a clinical conclusion between the device and the reported event.However, vessel characteristics and procedural factors may have contributed to the reported event.According to the safety information in the instructions for use ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM2CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
circuito int nte 1820
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key8470983
Report Number9616099-2019-02817
Device Sequence Number1664699
Product Code LIT
UDI-Device Identifier20705032059777
UDI-Public20705032059777
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model Number4400802S
Device Catalogue Number4400802S
Device Lot Number17731743
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2019
Supplement Date Received by Manufacturer04/16/2019
Initial Report FDA Received Date04/01/2019
Supplement Report FDA Received Date05/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-