• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R READY-SCREEN (2); REAGENT RED BLOOD CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R READY-SCREEN (2); REAGENT RED BLOOD CELLS Back to Search Results
Catalog Number 0006440
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On march 7, 2019 immucor technical support used a remote electronic connection method to review test records from the relevant run on the neo instruments serial number (b)(4).Review of the event log for both instruments show no errors during test event and review of the roi's for both test plates show that the roi's were not misaligned.On march 8, 2019 an immucor service technician inspected neo instrument serial number (b)(4) and performed an unexpected reaction checklist and the instrument was found to be operating within specifications.The internal immucor record for this event is (b)(4).
 
Event or Problem Description
On (b)(6) 2019 a customer reported unexpected negative screen results using capture-r ready-screen 2 on the neo instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-SCREEN (2)
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key8471164
Report Number1034569-2019-00066
Device Sequence Number16634912
Product Code KSZ
UDI-Device Identifier10888234000648
UDI-Public10888234000648
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
102707/5045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date03/20/2019
Device Catalogue Number0006440
Device Lot NumberX493
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2019
Initial Report FDA Received Date04/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-