Model Number 10-80-00 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Date of event is unknown.This information will be provided in a supplemental report if made available.Livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was not able to confirm the reported issue.The device was replaced and requested back to livanova deutschland for further investigation.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a s5 roller pump suddenly turned off during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
During the investigation at the manufacturer site, the reported malfunction could be reproduced.The pump turned off by slightly moving the connector of the cable set.The cause of the reported malfunction could be identified in a defective connector.The replacement of the cable solved the issue.
|
|