• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Loss of Power (1475)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Date of event is unknown.This information will be provided in a supplemental report if made available.Livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was not able to confirm the reported issue.The device was replaced and requested back to livanova deutschland for further investigation.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a s5 roller pump suddenly turned off during priming.There was no patient involvement.
 
Manufacturer Narrative
During the investigation at the manufacturer site, the reported malfunction could be reproduced.The pump turned off by slightly moving the connector of the cable set.The cause of the reported malfunction could be identified in a defective connector.The replacement of the cable solved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8472416
MDR Text Key140607417
Report Number9611109-2019-00216
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)160621
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-