| Catalog Number |
01L75-25 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information: all available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a false elevated architect cyclosporine result for a 69 year old male with myasthenia gravis.The sample generated a result of approximately 400 and retest generated 200.The customer indicated the patient's results prior to last month were less than 30.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of complaint activity did not identify any adverse or non-statistical trends for the architect cyclosporine assay.An increase in complaint activity was not identified for reagent lot 88026m800.Accuracy testing was also performed to evaluate the performance of reagent lot 88026m800.An internal panel was tested with retained kits of the likely cause reagent lot.Acceptance criteria was met, which indicates acceptable product performance.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.Additionally, labeling was reviewed and sufficiently addresses the customer's issue.No systemic issue or deficiency of the architect cyclosporine assay was identified.
|
| |
|