| Model Number |
98814-65 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 120 mg/dl and 400 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the xceed meter were reviewed and the dhrs showed the xceed meter passed all tests prior to release.Dhrs (device history review) for the precision strips were reviewed and the dhrs showed the precision strips passed all tests prior to release.Retain testing was performed for precision strips and all units performed in specification and passed.If the product is returned, the case will be re-opened, and a physical investigation will be performed.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 120 mg/dl and 400 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|