During preparation for the procedure, when the laser fiber was removed from the sterile packaging, it was noted the fiber was kinked/fractured.The device was set aside and a new of the same device was used to successfully complete the procedure.There was no harm or injury to the patient due to the event as the device did not come into contact with the patient.The disposable device has been returned to the manufacturer for a device evaluation.
|
|
Returned for evaluation was an ops evlt fiber.A visual evaluation of the disposable device noted the fiber tip end was fractured.The fracture occurred 8mm from fiber tip.The customer's reported complaint description of fiber fractured was confirmed.Review of the returned sample shows a fracture in the fiber.It is unknown how this kink/fracture occurred.A review of the lot history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.During the packaging step, the fibers are inspected for damage.A review of the angiodynamics' device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint # (b)(4).
|