• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Model Number LSMU1350716
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a manufacturing review could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent deployment procedure for treatment of cto in the innominate artery via femoral access, the balloon expandable vascular covered stent allegedly dislodged from the balloon prior to deployment.It was further reported that another balloon was used to post-dilate and apose the dislodged stent to the vessel wall.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT
Common Device Name
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
MDR Report Key8475044
Report Number9616666-2019-00036
Device Sequence Number16551409
Product Code PRL
UDI-Device Identifier05391522081543
UDI-Public(01)05391522081543
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P160024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSMU1350716
Device Catalogue NumberLSMU1350716
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2019
Initial Report FDA Received Date04/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-