The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was fractured into two pieces and both pieces were returned.There was significant damage to the device around the fracture site which could have contributed to the failure.The device was manufactured in 2015 and exhibits signs of extensive wear/usage.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|