• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Event or Problem Description
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's main keypad assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Physio-control further evaluated the removed main keypad assembly and determined that the cause of the reported issue was due to the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
 
Additional Manufacturer Narrative
This product complaint requires a supplemental mdr to document that field action number 301587611/18/2019-001c is relevant to the reported issue.
 
Event or Problem Description
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8476518
Report Number0003015876-2019-00573
Device Sequence Number13933310
Product Code MKJ
UDI-Device Identifier00883873911778
UDI-Public00883873911778
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001242
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/12/2019
Supplement Date Received by Manufacturer06/25/2019
11/18/2019
Initial Report FDA Received Date04/02/2019
Supplement Report FDA Received Date07/18/2019
11/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-11/18/2019-001C
Patient Sequence Number1
-
-