| Model Number |
15 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's main keypad assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Physio-control further evaluated the removed main keypad assembly and determined that the cause of the reported issue was due to the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
|
| |
|
Additional Manufacturer Narrative
|
|
This product complaint requires a supplemental mdr to document that field action number 301587611/18/2019-001c is relevant to the reported issue.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that when they attempt to deliver a test shock with the device they have to press the shock button several times.They also reported that the device has an event code logged in its memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
|
| |
|