• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 3.0X130MM FIBULA NAIL, LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 3.0X130MM FIBULA NAIL, LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3.0X130MM FIBULA NAIL, LEFT
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported during a left ankle fracture procedure, the proximal end of the fib30130l fibula nail broke off during insertion and was left inside of the fibula.The rep stated the fracture was successfully reduced and fixated with a distal fibula plate and screws.The seated depth of the implant was 130mm.A 3.22mm was used to prep the bone.Additional information has been requested.Additional information received on (b)(6) 2019: the rep stated that the surgeon attempted to remove the broken fragment, but no avail.The rep confirmed the remaining portion of the nail was retrieved and will be returning for evaluation.The bone was prepped by using a 6.2 distal reamer and 3.2 proximal reamer.The rep stated that the surgeon had to switch to their back up plan to make a larger incision so they could plate the fracture.A second fib30130l was not brought in.The case was completed by using an ankle fracture plate.
 
Additional Manufacturer Narrative
Complaint confirmed.The nose cone was found to have broken off from the device.The most likely cause is improper canal preparation, which caused the distal end of the device to become incarcerated into the unreamed canal and break when the implant was impacted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0X130MM FIBULA NAIL, LEFT
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key8479859
Report Number1220246-2019-00998
Device Sequence Number1650683
Product Code HSB
UDI-Device Identifier00816988011491
UDI-Public00816988011491
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K160069
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/01/2019
Device Model Number3.0X130MM FIBULA NAIL, LEFT
Device Catalogue NumberFIB30130L
Device Lot NumberA091117-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2019
Initial Date Received by Manufacturer 03/14/2019
Supplement Date Received by Manufacturer03/14/2019
Initial Report FDA Received Date04/03/2019
Supplement Report FDA Received Date04/09/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient Weight113
-
-