Device Problems Self-Activation or Keying (1557); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/04/2019
Event Type  malfunction  
Manufacturer Narrative
At the time of the reported event, user facility personnel realized the table was unlocked.A steris service technician arrived onsite to inspect the table and was unable to duplicate the reported event.No issue with the function or operation of the table was identified.Based on the description of the event and the technician's inspection, the cause of the reported event is user facility personnel inadvertently unlocking the table via the hand control which allows the table to move when pressure is applied (e.G, user facility personnel leaning on the table).A steris account manager offered in-service on the proper use and operation of the 4085 surgical table; however, the user facility declined.No additional issues have been reported.
 
Event Description
The user facility reported that at the beginning of a patient procedure, the 4085 surgical table moved without being commanded to do so.The procedure was completed successfully.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4085 SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key8479917
MDR Text Key141079048
Report Number1043572-2019-00025
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-