At the time of the reported event, user facility personnel realized the table was unlocked.A steris service technician arrived onsite to inspect the table and was unable to duplicate the reported event.No issue with the function or operation of the table was identified.Based on the description of the event and the technician's inspection, the cause of the reported event is user facility personnel inadvertently unlocking the table via the hand control which allows the table to move when pressure is applied (e.G, user facility personnel leaning on the table).A steris account manager offered in-service on the proper use and operation of the 4085 surgical table; however, the user facility declined.No additional issues have been reported.
|