TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The fsr replaced the central processing unit (cpu) board resolving the issue.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, an 'overspeed' error message was displayed.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed no overspeed errors to occur over the two days of testing with the central processing unit (cpu) board installed onto a lab use only test fixture.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|