AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR; DH EF PERC PLACEMENT PRODUCTS Back to Search Results
Model Number 98430
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 02 apr 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that after having an initial button placement in (b)(6) 2018, in (b)(6) 2019 a small metal rod from the suture button was removed from the site.It's possible to feel the same metal rod on the site of the two other sutures.No further information provided.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 22 aug 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
Two loose t-bar fasteners from the kit were received for evaluation.Visual examination revealed that there were no remnants of suture left on either fastener.There is no other damage to the t-bars.The complaint is confirmed as reported.No root cause could be determined.All information reasonably known as of 05 sep 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key8480820
Report Number9611594-2019-00063
Device Sequence Number1633845
Product Code KGC
UDI-Device Identifier10680651984309
UDI-Public10680651984309
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number98430
Device Catalogue Number991098430
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/20/2019
Supplement Date Received by Manufacturer08/01/2019
08/26/2019
Initial Report FDA Received Date04/03/2019
Supplement Report FDA Received Date08/23/2019
09/06/2019
Patient Sequence Number1
-
-