• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER; BARDEX LUBRISIL FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER; BARDEX LUBRISIL FOLEY CATHETER Back to Search Results
Model Number 175824
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Patient Problem/Medical Problem (2688); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the 24fr supra-pubic catheter would only go in about 2 inches and then stop.The patient reportedly went to the doctor where the md ran a rod into the opening in the stomach to remove what was blocking it and preventing the catheter from being inserted.The patient was reportedly switched to a 22fr catheter.
 
Event or Problem Description
It was reported that the 24fr supra-pubic catheter would only go in about 2 inches and then stop.The patient reportedly went to the doctor where the md ran a rod into the opening in the stomach to remove what was blocking it and preventing the catheter from being inserted.The patient was reportedly switched to a 22fr catheter.Additional information was received from the complainant via phone on 20-mar-2019, that his stoma was not newly placed and he had been using a 24 fr catheter for 2 years.There appeared to be something blocking the inside of his stoma that would not allow the catheter to be fully inserted.He stated that the doctor placed a scope in his bladder to open it up.He was now using a 22 fr catheter and had not had any further issues.
 
Additional Manufacturer Narrative
This report is being submitted past the regulatory timeframe.Please see attached letter for additional information.The device was not returned for evaluation.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Unless package is opened or damaged.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.Note: aggressive traction, particularly in the presence of suturing, is not recommended for silicone foley catheters.Bard, bardex and lubri-sil are registered trademarks of c.R.Bard, inc.Or an affiliate.U.S.Patent number 5,179,174 and patent pending.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." patient code: 2519 h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.- attachment: [fda letter for late mdrs.Pdf] h3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
BARDEX LUBRISIL FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8481056
Report Number1018233-2019-01719
Device Sequence Number13820992
Product Code EZL
UDI-Device Identifier00801741034206
UDI-Public(01)00801741034206
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number175824
Device Catalogue Number175824
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/11/2019
Supplement Date Received by Manufacturer04/08/2019
Initial Report FDA Received Date04/03/2019
Supplement Report FDA Received Date05/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-