The pusher was found broken and kinked distal to the break indicator with the release wire extended out the proximal end; along with the actuator interface, coupler tube, positive load indicator, and break indicator missing and not returned.It appeared that manual detachment was attempted at this location.In addition, the axium prime pusher was broken at ~23.0cm from the proximal end but is retained by the release wire.Kinks and bends were found along the length of the pusher between ~23.0cm to ~143.5cm from the proximal end.The coin was not present at the lumen stop as it was pulled back.The shield coil is present and intact.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations and was found to be within specifications.Measured 0.078mm @ 0.063mm; measured 0.083mm @ 0.127mm; measured 0.096mm @ 0.275mm.The inner diameter of the lumen stop and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00288¿ and found to be within specification (0.00220" minimum - 0.0029" maximum).The retainer ring inner diameter (id) was measured to be 0.00456¿and found to be within specification (0.00455"+/- 0.00010").The introducer sheath was found to be bent at ~13.0cm and ~57.0cm from the proximal end.No other anomalies were observed.Based on the analysis performed the axium prime coil was confirmed to have prematurely detached.There was no malfunction of the pusher or non-conformance to specification that may have contributed to the pre-mature detachment.Since the actuator interface, coupler tube, positive load indicator, break indicator, implant coil and detached element, any contribution of the actuator interface, coupler tube, positive load indicator, break indicator, implant coil and detached element to the issue could not be determined.Based on the event description it is was not possible to ascertain the cause of the pre-mature detachment; however, based on the returned device, there was evidence of manual detachment attempts to detach the coil; with the coin pulled back from the lumen stop.The pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.Furthermore, the pusher was bent and kinked along its length, indicative of high tortuosity or damage during return shipping to medtronic for analysis.The introducer sheath was found to be bent, however, as there is no physician report of damage to the introducer sheath, and the bends corresponds to the same bent sections of the pusher, it is likely that the bends occurred during shipping of the device back to medtronic for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|