Model Number IS20-25/C65 SA
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx abdominal aortic aneurysm stent.Approximately three (3) years post initial procedure, the patient presented with a "disjunction of the iliac extension", characterized as a type iiia endoleak with component separation.The physician elected to treat the patient by relining with two (2) additional limbs stent afx grafts on (b)(6) 2019.There have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
A clinical evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and or imaging; requests were made and no response was received.As such, event determination, off-label conditions, related patient harms, and patient disposition could not be independently assessed.The manufacturing lot review confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
LIMB STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8484535
Report Number2031527-2019-00165
Device Sequence Number1575852
Product Code MIH
UDI-Device Identifier00818009011473
UDI-Public(01)00818009011473(17)180213
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 03/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2018
Device Model NumberIS20-25/C65 SA
Device Lot Number1317722-022
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2019
Supplement Date Received by Manufacturer03/08/2019
Initial Report FDA Received Date04/04/2019
Supplement Report FDA Received Date05/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0009-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-BIFURCATED STENT GRAFT:1251893-015; AFX-INFRARENAL AORTIC EXTENSION:1404178-005; AFX-LIMB STENT GRAFT:1252621-011; AFX-BIFURCATED STENT GRAFT:1251893-015; AFX-INFRARENAL AORTIC EXTENSION:1404178-005; AFX-LIMB STENT GRAFT:1252621-011
Outcome Attributed to Adverse Event Required Intervention;
-
-