MAUDE Adverse Event Report: DEPUY MITEK LLC US GII/SUPER/RC DRGDE SLOT *EA; SURGICAL HAND DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US GII/SUPER/RC DRGDE SLOT *EA; SURGICAL HAND DRILL Back to Search Results
Catalog Number 213207
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The lot number is unknown.
 
Event or Problem Description
It was reported by the sales rep via phone that during a posterior tibial tendon repair procedure the gii slotted drill guide created metal shavings as the drill bit was being passed through.The sales rep stated that it seemed the guide was bent.The shavings were removed with tweezers and then flushed with irrigation tubing, so that no debris remained in the patient.The bone hole originally drilled for the case was able to be used.The case was completed with this device with no patient harm or delay.The sales rep could not provide a lot number.The device is being returned for evaluation.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Further, no lot numbers were supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SURGICAL HAND DRILL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8484616
Report Number1221934-2019-56769
Device Sequence Number1471652
Product Code FZX
UDI-Device Identifier10886705002610
UDI-Public10886705002610
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number213207
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/12/2019
Supplement Date Received by Manufacturer05/15/2019
Initial Report FDA Received Date04/04/2019
Supplement Report FDA Received Date05/16/2019
Patient Sequence Number1
-
-