• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X 60MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X 60MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, POROUS Back to Search Results
Catalog Number OR71332760
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 10/19/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that right hip revision surgery was performed due to recurrent dislocations.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.The clinical/medical team concluded, the clinical information provided, (b)(6)) the elevated metal ion levels, synovitis and staining may be consistent with a reaction to metal debris.However the source cannot be determined with the available documentation.Also, it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The components revised were the bhr resurfacing, the r3 liner and r# shell.In conclusion, for (b)(6) the root cause of the dislocations cannot be concluded, however the mixture of components during the 1st right revision with different manufacturers could have contributed to the issue.It is noted that no x-rays copies or components were returned for this investigation.Patient impact except for the surgeries cannot be predicted and a follow-up status has not been provided at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual device involved our investigation cannot proceed.If the device or new information is received in the future, the complaints can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 36MM X 60MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8485195
Report Number1020279-2019-01394
Device Sequence Number11208552
Product Code MBL
UDI-Device Identifier03596010598028
UDI-Public03596010598028
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberOR71332760
Device Lot Number16MM01995
Initial Date Received by Manufacturer 03/19/2019
Supplement Date Received by Manufacturer03/19/2019
Initial Report FDA Received Date04/04/2019
Supplement Report FDA Received Date11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
-
-