One sample was received dry, unused and in an open pouch.A visual inspection with naked eye noted a scratch or gouge on the patient line tubing approximately 8¿ from the patient connector.The device did not leak as the scratch or gouge did not penetrate the tubing wall.Functional testing, including leak testing, clear passage testing, and clamp function testing were performed with no issues noted.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|