| Model Number |
5810 |
| Medical Device Problem Codes |
Difficult to Interrogate (1331); Pacemaker Found in Back-Up Mode (1440)
|
| Health Effect - Clinical Codes |
Muscle Stimulation (1412); Discomfort (2330)
|
| Date of Event |
03/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the device exhibited backup vvi operation most probably due to low battery.Due to poor telemetry the device was not restored from backup to normal functions.The patient experienced pocket stimulation and discomfort.No intervention was performed.The patient will continue to be monitored.
|
| |
|
Event or Problem Description
|
|
New information states that device replacement was discussed.The patient was in a stable condition.
|
| |
|