• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINNACLE DP REV ACET CUP 66MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINNACLE DP REV ACET CUP 66MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121707066
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Scar Tissue (2060); Tissue Damage (2104); Discomfort (2330); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 12/08/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Initial reporter occupation: non-healthcare professional: "attorney".(b)(4).
 
Event or Problem Description
Litigation alleges the patient suffers from pain, discomfort, and toxic cobalt-chromium metal ions and particles to be released into the blood, tissue and bone.Update: sales rep reported revision surgery.Patient was revised to address instability.Update: in addition to what were previously alleged, ppf alleges loosening of cup, metal wear, metallosis and elevated metal ions.Update: in addition to what were previously alleged, pfs alleges that the patient felt like there was scraping sensation in his hip, walking difficulty and falling down on occasion.Surgeon recommended revision surgery due to loose component and metallosis.After review of medical records, the patient was revised to address cup loosening, metallosis and instability.Revision notes reported that hip was found to be dislocated, synovial scarring, and grossly loose cup.It was also found that the stem is 20 degrees retroverted.The sleeve was left in place and stem was rotated over 40 degrees to provide appropriate anteversion then was re-inserted.Pathology report is positive for acute inflammation.Lab data shows metal ion levels were above 7 ppb.Doi: (b)(6) 2007; dor: (b)(6) 2014; (right hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE DP REV ACET CUP 66MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
MDR Report Key8486596
Report Number1818910-2019-89878
Device Sequence Number14121216
Product Code LPH
UDI-Device Identifier10603295008934
UDI-Public10603295008934
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K033338
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2014
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2009
Device Catalogue Number121707066
Device Lot NumberYS6FJ1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/11/2019
Supplement Date Received by Manufacturer12/03/2019
Initial Report FDA Received Date04/05/2019
Supplement Report FDA Received Date12/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight113
-
-