| Catalog Number |
18915030S |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
When nurse opened the sterile screw box, she saw a little hair (looks like an eye lash or eyebrow).Surgery was delayed a little - according to details received, less than 15 min.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the device was unpacked.The device inspection revealed the following: we received the locking screw, without overwrap-foil, in its original packaging.After unpacking the blister becomes visible with a completely pulled off tyvek® lid.Inspection of the tyvek® lid revealed a hair [like an eye-lash] at the bottom site at the area of the rim, which could be easily captured.The locking screw shows no traces of use.The reported event could not be reproduced as the packaging was received with an already broken sealing.Pollution in original condition could not be verified.Due to adhesive power of the glued / sealed connection of tyvek - blister a white colored area remains on the blister when the tyvek® lid is torn off.This white area is called crazing or stress-whitening.A consistent white and not interrupted surface indicates that no sealed-in foreign material (e.G.Hair) at the sealed area of the blister rim was present.In case of a sealed in hair a negative imprint would be apparent at the stress-whitening.Furthermore, not enough information is available to confirm the reported case.No deviations were reported during in-control steps performed.Finally, it could not be excluded that the pollution occurred at the user site.Based on the above observations the root cause could not be determined.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.
|
| |
|
Event or Problem Description
|
|
When nurse opened the sterile screw box, en she saw a little hair( looks like a eye lash or eyebrow).Surgery was delayed a little - according to details received less than 15 min.
|
| |
|